User Manual

Command Descriptions
Group
Power and Batte
ry-Related
Syntax
POWer?
Examples
POWER? might r eturn the following: POWER:AC:PRESENT
1;:POWER:BATTERY1:GASGAUG E 0 ;STATUS
NOTINSTALL
ED;:POWER:BATTERY2:GASGAUGE 0 ;STATUS
NOTINSTALLED;:POWER:BATTE RIES:TIME 0; POWER:BUTTON LIGHT 1
POWer:AC:PRESENt? (Query Only
)
Returns a statement indicating whether the o scilloscope is being powered by
battery or AC.
Group
Power and Battery-Related
Syntax
POWer:AC:PRESENt?
Returns
0 if the oscilloscope is being powered by battery. 1 if the oscilloscope is being
powered by AC.
Examples
POWER:AC:PRESENT? might return 0, specifying that the oscilloscope is being
powered by its battery.
POWer:BATTERY<x>:GASgauge? (Query Only)
Returns the amount of charge remaining in B
attery x, where x is Battery 1 or
Battery 2.
This command is valid only if POWer:BATTERY<x>:STATUS? returns
NORMAL.
Group
Power and Battery-Related
Syntax
POWer:BATTERY<x>:GASgauge ?
Returns
<NR1>, a number between 0 and 100, indicates the charge remaining in the
battery. 0 indicates the battery is fully discharged. 100 indicates the battery
is fully charged.
2-134 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer