Command Descriptions
MAXImum (TDS10
00, TDS2000, TDS1000B, TDS2000B, TDS2000C, and
TPS2000 Series only) is the value of the largest point in the waveform.
MEAN is the arithmetic mean over the entire waveform.
MINImum (TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and
TPS2000 Series only) is the value of the smallest point in the waveform.
NONE disables the measurement specied by <x>.
NWIdth (TDS200 Series with a TDS2MM measurement module, TDS1000,
TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series only) is the
negative pulse width between the rst falling edge and the next rising edge at the
waveform 50% level. Falling and rising edges must be displayed to measure. The
oscilloscope automatically calculates the 50% measurement point.
PDUty (TDS2000C Series) is the ratio of the pulse width to the signal period,
expressed as a percentage.
PERIod is the duration, in seconds, of the rst complete cycle in the waveform.
PHAse (TDS2000C Series) is the phase difference from the se l ected waveform
to the designated waveform.
PK2pk is the absolute difference between the maximum and m inimum amplitude.
PWIdth (TDS200 Series with a TDS2MM measurement module, TDS1000,
TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series only) is the
positive pulse width b etween the rst rising edge and the next falling edge at the
waveform 50% level. Rising and falling edges must be displayed to measure. The
oscilloscope automatically calculates the 50% measurement point.
RISe (TDS200 Series with a TDS2MM measurement module, TDS1000,
TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series only) is the
rise time between 10% and 90% of the rst rising edge of the waveform. Rising
edge must be displayed to meas
ure. The oscilloscope automatically calculates
the 10% and 90% measurement points.
Examples
MEASUREMENT:MEAS3:TYPE CR
MS
speciesMEAS3tocalculatethetrueRoot
Mean Square value.
MEASUrement:MEAS<x>:UNIts? (Query Only)
Returns the units for the oscilloscope measurement specied by
MEASUrement:MEAS<x>:TYPe. Where <x> identies the measurement, 1
through 4 for the TDS200 Series, and 1 through 5 for the TDS1000, TDS2000,
TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series.
Group
Measurement
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-127