User Manual

Command Descriptions
Arguments
CH<y> species
the input channel source for the measurement.
MATH species the measurement source channel as the math waveform.
(TDS1000B, TD
S2000B, TDS2000C, and TPS2000 Series only)
Examples
MEASUREMENT:MEAS2:SOURCE CH1 sets the source for Measurement 2 to
channel 1.
MEASUrement:MEAS<x>:TYPe
Sets or queries the on-screen periodic oscilloscope measurement type for the
measurement specied by <x>. Where <x> identies the measurement, 1
through 4 for the TDS200 Series, and 1 through 5 for the TDS1000, TDS2000,
TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series.
This is equivalent to selecting the measurement type in the MEASURE menu.
Setting the type to anything other than NONE d isplays the MEASURE menu
on the screen.
NOTE
. You should use the MEASUrement:IMMed command with programming to
takemeasurements,asthisispreferredto the MEASUrement:MEAS<x>command.
Gro
up
Mea
surement
S
yntax
MEASUrement:MEAS<x>:TYPe { CRMs | CURSORRms | DELay | FALL |
FREQuency | MAXImum | MEAN | MINImum | NONe | NWIdth | PDUty
| PERIo d | PHAse | PK2pk | PWIdth| RISe }
MEASUrement:MEAS<x>:TYPe?
Arguments
CRMs is the true Root Mean Square voltage of the rst complete cycle in the
waveform.
CURSORRms (TDS2000C Series) is the true Root Mean Square (RMS) value of the
selected waveform data, between the start point and the end point.
DELay (TDS2000C Series) is the delay from one waveform’s edge event to
another.
FALL (TDS200 Series with a TDS2MM measurement module, TDS1000,
TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series only) is the
fall time between 90% and 10% of the rst falling edge of the waveform. Falling
edge must be displayed to measure. The oscilloscope automatically calculates
the 10% and 90% measurement points.
FREQuency is the reciprocal of the period measured in Hertz.
2-126 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer