User Manual
Command Descriptions
Examples
MEASUREMENT:I
MMED:VALUE?
might return 28.75E6 if you are measuring
the frequency of a 28.75 MHz signal.
MEASUREMENT:
IMMED:VALUE?
might return 9.9E37, which means that the
channel specified by MEASUrement:IMMed:SOUrce is not currently displayed.
*ESR? after
a measurement query might return
16, indicating that there are error
messages in the queue.
ALLEV might
return
ALLEV 5 41,"Measurement warning, Low signal
amplitude: "
MEASUrement:MEAS<x>? (Query Only)
Returns all measurement parameters for the displayed o scilloscope periodic
measurement specified by <x>. Where <x> identifies the measurement, 1
through 4 for the TDS200 Series, and 1 through 5 for the TDS1000, TDS2000,
TDS100
0B, TDS2000B, TDS2000C, and TPS2000 Series.
Group
Measurement
Syntax
MEASUrement:MEAS<x>?
Returns
Settings for the specifie d measurement source.
Examples
MEASUREMENT:MEAS3? might return PERIOD;"s";CH1
MEASUrement:MEAS<x>:SOUrce
Sets or queries the source for an automated measurement. Where <x> identifies
the measurement, 1 through 4 for the TDS200 Series, and 1 through 5 for the
TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series.
This is equivalent to selecting the measurement source in the MEASURE menu.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce { CH<y> | MATH }
MEASUrement:MEAS<x>:SOUrce ?
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-125