User Manual

Command Descriptions
PFPHASE (TPS20
00 Series with TPS2PWR1 Power Analysis Module only) is the
phase angle measurement in degrees. Source 1 must be v olts; source 2 must
be Amperes.
PHAse (TPS2000 Series with TPS2PWR1 Power Analysis Module only) is the
measurement of phase betwee n source 1 and source 2, in degrees.
POWERFACTOR (TPS2000 Series with TPS2PWR1 Power Analysis Module only)
is the true power factor ratio measurement. Source 1 must be volts; source 2 must
be Amperes.
TRUEPOWER (TPS2000 Series with TPS2PWR1 Power Analysis Module only)
is the true power measurement in Watts. Source 1 must be volts; source 2 must
be Amperes.
VAR (TPS2000 Series with TPS2PWR1 Power Analysis Module only) is the
reactive power measurement. Source 1 must be volts; source 2 must b e Amperes.
WFCREST (TPS2000 Series with TPS2PWR1 Power Analysis Module only) is the
measurement of the maximum value to the cycle RMS value of the waveform,
a unit-less ratio.
WFCYCRMS (TPS2000 Series with TPS2PWR1 Power Analysis Module only)
is the measurement of RMS voltage calculated over the rst cycle, using the
min-m
ax, high-low method.
WFFREQ (TPS2000 Series with TPS2PWR1 Power Analysis Module only) is the
meas
urement of frequency using the min-max, high-low method.
Examples
MEASUREMENT:IMMED:TYPE FREQUENCY denes the immediate meas urement to
be
a frequency measurement.
MEASUrement:IMMed:UNIts? (Query Only)
Returns the units for the immediate oscilloscope measurement.
Group
Measurement
Syntax
MEASUrement:IMMed:UNIts?
Returns
<QString> returns "V" for volts, "s" for seconds, or "Hz" for Hertz.
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-123