User Manual

Command Descriptions
Conditions
This command ap
plies to the TPS2000 Series with TPS2PWR1 Power Analysis
Module only.
NOTE. This command is only available when the Power Analysis Module
application k ey is installed.
Group
Measurement
Syntax
MEASUrement:IMMed:SOURCE2 { CH<x> | MATH }
MEASUrement:IMMed:SOURCE2?
Arguments
CH<x> species the measurement source channel a s one of the input channels. The
value of <x> can vary from 1 through 4 for 4-channel instruments or 1 through 2
for 2-channel instruments. For power analysis, Source2 must be Amperes.
MATH species the measurement source channel a s the math waveform.
Examples
MEASUREMENT:IMMED:SOURCE2 CH2 species channel 2 as the immediate
measurement source.
MEASUrement:IMMed:TYPe
Sets or queries the immediate measurement type.
NOTE. Selecting one of the power ar
guments for this command, without installing
the Power Analysis Module application key, will generate an error.
Group
Measurement
Syntax
MEASUrement:IMMed:TYPe { CRMs | CURSORRms | DELay | FALL |
FREQuency | MAXImum | MEAN | MINImum | NONe | NWIdth | PDUty
|PERIod | PHAse | PK2pk | PWIdth| RISe }
TPS2000 with a TPS2PWR1 Power Analysis Module only:
{ WFCRE ST | WFFREQ | WFCYCRMS | TRUEPOWER | VAR |
POWERFACTOR | PFPHASE | Phase }
MEASUrement:IMMed:TYPe?
Arguments
CRMs is the true Root Mean Square voltage of the rst complete cycle in the
waveform.
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-121