User Manual

Command Descriptions
MEASUrement:
IMMed? (Query Only)
Returns all immediate measurement setup parameters. Immediate queries
and commands are the preferred methods for programming. A n immediate
measurement
selection is not visible or accessible through the display screen
or front panel.
Group
Measurement
Syntax
MEASUrement:IMMed?
Returns
Immediate measurement setup parameters
Examples
MEASUREMENT:IMMED? might return PERIOD;"s";C H1
MEASUrement:IMMed:SOUrce[1]
Sets or queries the source for single-sourc e immediate measureme nts .
Group
Measurement
Syntax
MEASUrement:IMMed:SOUrce[ 1] { CH<x> | MATH }
MEASUrement:IMMed:SOUrce[ 1]?
Arguments
CH<x> species the measurement source channel as one of the input channels.
The value of <x> can vary from 1 through 4 for 4-channel instruments or 1
through 2 for 2-channel instruments.
MATH species the measurement source channel as the math waveform.
(TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series only)
Examples
MEASUREMENT:IMMED:SOURCE CH1 species channel 1 as the immediate
measurement source.
MEASUrement:IMMed:SOURCE2
Sets or queries the secondary source for dual-source immediate measurements.
For example, power analysis and phase angle measurements.
2-120 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer