User Manual
Command Descriptions
LIMit:TEMPLat
e:TOLerance:VERTical to create a limit test template and
store the template in the reference waveform memory location specified in
LIMit:TEMPLate:DESTination. You can compare the waveform acquired through
the specified channel against this template. The template can be a waveform saved
in any of the reference locations REF1 through REF4.
Group
Limit Test
Syntax
LIMit:TEMPLate APPLY
Related Commands
LIMit:TEMPLate:DESTination, LIMit:TEMPLate:SOUrce, LIMit:TEMPLate:
TOLerance:HORizontal, LIMit:TEMPLate:TOLerance:VERTical
Arguments
APPLY creates a templa te with the specified source waveform and tolerances,
and stores it in the destination reference waveform to be used in the limit test
comparisons.
Examples
LIMIT:TEMPLATE APPLY
creates a template with the specified source waveform and tolerances, and stores it
in the destination reference waveform to be used in the limit test comparisons.
LIMit:TEMPLate:DESTination
Sets or returns the destination reference waveform that the LIMit:TEMPLate
APPLY command will use.
Group
Limit Test
Syntax
LIMit:TEMPLate:DESTinatio n REF<x>
LIMit:TEMPLate:DESTinatio n?
Related Commands
LIMit:TEMPLate, LIMit:TEMPLate:SOUrce, LIMit:TEMPLate:TOLerance:
HORizontal, LIMit:TEMPLate:TOLerance:VERTical
Arguments
REF<x> spec ifies the reference waveform destination in which the template
waveform is to be stored.
2-11 0 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer