User Manual

Command Descriptions
Syntax
LIMit:STOPAft
er:MODe
{ MANua l | WAVEform | VIOLation | TIMe }
LIMit:STOPAfter:MODe?
Related Commands
LIMit:STOPAfter:TIMe, LIMit:STOPAfter:VIOLation, LIMit:STOPAfter:
WAVEform
Arguments
MANual causes the test to stop after an operator manually stops it.
WAVEform causes the test to stop after the oscilloscope tests a s
pecied number
of waveforms.
VIOLation causes the test to stop after the oscilloscope detects a v iolation.
TIMe causes the test to stop after the oscilloscope detects that the preset running
time has elapsed.
Examples
LIMIT:STOPAFTER:MODE?
might return
VIOLATION
LIMit:STOPAfter:TIMe
Sets or returns the limit test stop after time setting. The limit test will automatically
stop when it reaches the specied time, in seconds.
Group
Limit Test
Syntax
LIMit:STOPAfter:TIMe <NR3>
LIMit:STOPAfter:TIMe?
Related Commands
LIMit:STOPAfter:MODe, LIMit:STOPAfter:VIOLation, LIMit:STO PAfter:
WAVEform
Arguments
<NR3> is the time in seconds to run the limit test.
Examples
LIMIT:STOPAFTER:TIME 10
sets the limit test to run for 10 seconds.
2-108 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer