User Manual

Command Descriptions
Related Commands
LIMit:SAVEWFM
Arguments
OFF or <NR1> = 0 turns off the function to save waveform image data to the ash
drive for limit testing.
ON or <NR1> 0 turns on the function to save waveform image data to the ash
drive for limit testing.
Examples
LIMIT:SAVEIMAGE ON
turns on the function to save waveform image data to the ash drive for limit
testing.
LIMit:SAVEWFM
Sets or returns whether the oscilloscope should save the waveform to a le on a
limit test failure event. When enabled, this command copies wav
eform data from
all active waveforms that have limit test failures to a le on the oscilloscope
ashmemorydrive.
Group
Limit Test
Syntax
LIMit:SAVEWFM { OFF | ON | 0 | 1 }
LIMit:SAVEWFM?
Related Commands
LIMit:SAVEIMAge
Arguments
OFFor<NR1>=0turnsoffthe
function to save waveform data to the ash
drive for limit testing.
ON or <NR1> 0 turns on the fu
nction to save waveform data to the ash drive
for limit testing.
Examples
LIMIT:SAVEWFM OFF
turns off the feature to save the waveform to a le on a limit test failure.
LIMit:SOUrce
Sets or returns the source channel for the wavefo
rm currently being compared to
the test template.