User Manual
Command Descriptions
Arguments
<NR3> is the pos
ition in seconds. This value is the difference between the trigger
point and the center graticule. Positive values place the trigger before the center
graticule.
Examples
HORIZONTAL:POSITION 2.0E-6 sets the main trigger position to 2ms before the
center graticule.
HORIZONTAL:POSITION? might return -1.0 E-3 , indicating that the main trigger
position is 1 ms after the center graticule.
HORizontal:RECOrdlength? (Query Only)
Returns the number of acquisition data points. For TDS200, TDS1000, TDS2000,
TDS1
000B, TDS2000B, TDS2000C, and TPS2000 oscilloscopes, this value is
always 2500, even in FFT mode. This command is provided to maintain program
compatibility with other Tektronix digital oscilloscopes.
Group
Horizontal
Syntax
HORizontal:RECOrdlength?
Related Commands
WFMPre:NR_Pt?
Returns
2500
Examples
HORIZONTAL:RECORDLENGTH? returns 2500 as the number of data points in
each record.
HORizontal:SCAle
Sets the time per division for the main time base and is identical to the
HORizontal:MAIn:SCAle command. It is included for compatibility purposes.
Group
Horizontal
Syntax
HORizontal:SCAle
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-99