Manual

Specications
These specications are valid only under the following conditions:
The probe and amplier have been calibrated at an ambient temperature of
23 ° ±5 °C.
The probe and amplier are operating within the environmental limits
described in Table 4-6 on page 4-5 (the operating temperature limits are 0 °C
to +50 °C, unless otherwise stated).
The probe and amplier have had a warm-up period of at least 20 minutes.
The probe degauss/autobalance routine has been performed after the
20-minute warm-up period, and thereafter whenever the PROBE
DEGAUSS/AUTOBALANCE light blinks.
The amplier output is properly terminated into 50 .
Specications are separated into two categories: warranted specications and
nominal or typical characteristics.
Warra
nted Specications
Warranted spec ications, Table 4-1, are guaranteed performance specications
unless specically designated as typical or nominal.
Tabl
e 7: Warranted TCPA300 and TCPA400 specications
Amplier TCPA300 TCPA400
DC Gain Accuracy 1% 1%
Installed probe TCP312 TCP305 TCP303 TCP404XL
Bandwidth (-3 dB) DC to 100 MHz DC to 50 MHz DC to 15 MHz DC to 2 MHz
Rise Time, 10% to
90%
3
.5 ns
7ns
2
3ns
1
75 ns
DC Gain Accuracy:
Warranted
3% 3% 3%
1
3%
Typical
2
1% 1% 1% 1%
1
Warranted from 10 °C to 50 °C. For temperature range of 0 °C to <10 °C, spec is +3%/-6%.
2
23 °C ±5 °C
TCPA300/400 Ampliers and TCP300/400 Series Current Probes User Manual 47