Troubleshooting and Error Codes
Displaying Er
ror Codes with the Probe Degauss Autobalance Button
This section describes the error codes that the amplifiers display using the
function indicator LEDs.
When an internal error c ondition exists, the amplifiers may generate error codes.
To display the error codes, do the following. (See Figure 25.)
1. Press the PROBE DEGAUSS AUTOBALANCE button.
2. The AC and DC Coupling LEDs will flash alternately to indicate that error
codes are being displayed instead of normal conditions.
3. The four fault indicator LEDs above the ON/STANDBY switch a re used to
form a four-bit binary error code. (See Figure 26 on page 44.) The error codes
are defined in the table. (See Table 6 on page 44.)
The PROBE OPEN LED indicates the most significant bit (MSB), and the
NONCOMPATIBLE PROBE TYPE LED indicates the least significant bit
(LSB).
4. To continue past an error code , press any button except ON/STANDBY.
Howev
er, the degauss will fail until the internal error c ondition is corrected
and the degauss operation is run again.
Figure 25: Error code display
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes User Manual 43