Manual

Application Notes
Inductance Measurements
You can us e the TCPA300 and TCPA400 to measure inductance of coils. Two
different methods can be used: one for low-impeda nce pulse sources and another
for high-impedance pulse sources of known value.
Low-Impedance Pulse
Sources
This gure shows a constant-voltage pulse generator of extremely low output
impedance connected to an inductor that has low resistance. (See Figure 20.)
1. Connect the inductor across the output terminals of the pulse generator.
2. Maintain a constant voltage across the inductor.
3. Clamp the
current probe over one of the source leads.
Figure 20: Measuring inductance with a low-impedance source
NOT
E. Iftheprobeimpedanceisasignicant part of the total circuit inductance,
measurement accuracy will be affected. Refer to the probe specications for
probe insertion impedance.
TCPA300/400 Ampliers and TCP300/400 Series Current Probes User Manual 35