Disassembly
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
7-5
3. Remove the 6 screws that secure the right-side inner panel to the chassis. See
Figure 7--2.
Right--side panel
Amplifier
Screws (6)
Torque screws to 8 in-lbs
when reassembling
Adjustment
access ports