1. Before reassembling the pro be body, be sure to verify the following:
a. The gap between the stationary and moveable core pieces is clean. If necessary, use isopropyl alcohol or a similar
cleaning agent to clean the pieces.
b. The contacts of the slide switch are clean. (Clean them if nece ssary.)
c. The plastic slide assembly is lubricated. If necessary, apply silicone-based grease sparingly to the parts.
2. Reassemble the probe body by following steps 1 and 2 of the probe head disassembly procedure in reverse. (See page 71,
Probe Disassembly.)
CAUTION. To avoid damaging the wires inside the probe, carefully dress the wires in the lower body half to avoid contact with
the slide assembly.
To avoid damaging the switch contacts, exercise care when tting the slide back into the probe body. Aligning the switch
contacts requires precision.
If the slider does not work smoothly after tightening the two screws, loosen the screws slightly.