Adjustments
Adjust the 150 A Range
6. Set the prob e range to 150 A.
7. Set the calib rator output to 7.5 A.
8. Adjust the 150 A gain on the probe to
display 750 mV, +/- 3 mV on the DMM.
9. Disconnect the pro be from the current
source.
TCP0150 C urren t Probe Instruction Manual 69