Adjustments
Adjustments
The pro cedures that follow de scribe the DC gain accuracy adjustments to the p robe to bring the performance within the warranted
specications.
Equipment Required
Refer to the Perf ormance Verication procedure for the equipment required. You also need an insulated ad justment tool. (See
page 28, Adjustment Tool.)
Equipment Setup
Refer to the Performance Verication procedure for the equipment setup.
CAUTION. To avoid ESD da mage to the probe, use an a ntistatic wrist strap and work at a static-approved workstation when
handling the probe.