Performance Verification
Equipment Setup
Use the following procedure to set up and warm up the equipment to test the probe.
1. Turn on the oscillosco pe.
2. Connect the probe to any chan nel (1–4) of
the oscilloscope.
3. Press the Degauss/Au toZero bu tton.
4. Power on the DMM, current source, and
the pulse generator.
5. Allow 20 minutes for the equ ipment to
warm up.
6. Photocopy the test record and use it to
record the test results. (See page 66.)
60 TCP0150 Current Probe Instruction Manual