Performance Verification
Equipment Required
Table 6 lists the e quipmen t required for the performance veri fication procedure.
Table 6: Test equipment
Description and quantity Performance requirement Recommended ex ample
1
Oscilloscope TekVPI interface, 500 MHz or gre ater bandwidth Tektronix D PO40 00
High Amplitude Pulse
Generator
Risetime <1 ns, pulse width >100 ns, amplitude
>10 Vpp into 50 Ω
Picosecond Labs 2600C
Calibrator DCA: 0.25% accuracy, 0 to ±10 A
ACA: 0.25% accuracy, 0 to ±7.5 A, square wave
output
Fluke/Wavetek 9100 w/Options 100
and 250 or 600
Digital Multimet er (DMM) DCV: 0.2% accuracy
Keithley 2700
TekVPI Calibration/Verification
adapter
TekVPI interface
067-1701-XX
DC Current loop 5 turns 18 AWG coa ted wire on 3 inch form See instructions that follow
HF Current loop 50 Ω ±0.5%, BNC male
015-0601-50
BNC-to-Dual banana adapter
103-0090-00
SMA M-to-BNC F adapter
015-0554-00
BNC Cable 50 Ω , 0.76 m (30 in) length
012-0117-00
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers .
58 TCP0150 Current Probe Instruction Manual