User Manual

Accessories and Options
Deskew/Calibration Fixture
Connect this xture to host instruments that
support the probe calibration or deskew
procedures. The deskew proced ures
compensate for gain errors and timing
differences between current and voltage
probes. Refer to your oscilloscope manual or
xture documentation for instructions.
WARNING. To reduce the risk of shock, do not
use this accessory on voltages above 30 Vrms,
42 Vpk, or 60 VDC.
This accessory is limited to 2 Wrms an d is not
intended to be used with hazardous volta ges.
Reorder Tektronix part number: 067-1686-00
TCP0150 C urren t Probe Instruction Manual 27