User Manual
Installation
Degauss/AutoZero
Button and LED
When the multicolor Degauss/AutoZero status
LED fla shes red, you must degauss the probe.
If the LED flashes orange, you should degauss
the probe. The DC gain and offset accuracy are
not guaranteed when this LED flashes orange.
The Degauss/AutoZero function also clears
(AutoZeroes) any DC offset in the probe.
To degauss the probe , do the following:
1. Disconnect the probe from the current
source and lock the slide.
2. Press the Degauss/AutoZ ero button to
initiate the degauss routine.
The LED glows green after successfully
completing the Degauss/AutoZero routine.
12 TCP0150 Current Probe Instruction Manual