d. Set the host name of the TBX3000A to correspond to the bench on which it is
located.
e. You should change the default password of the TBX3000A to avoid the settings
being changed by other users.
3. Install and configure the lab server software. Please refer to Installing TSL3000B for
more details, including installing and configuring the TSL3000B application software,
.net framework, Web Service (IIS6.0 or higher) and SQL Server 2008 R2 on the lab
server.
Note: The Enable Database & Webserver should not be selected if Web Service (IIS6.0
or higher) and SQL Server 2008 R2 have not been installed.
4. Connect the lab server to the WIFI network you set up previously, and set the WIFI
IP to use a static IP (for example 192.168.1.66). The static IP must not be within the
DHCP pool of the WIFI network.
Note: When the lab server is in DHCP mode, the IP address could be changed after
recycling, in which case the web page address of each bench will be changed
accordingly.
5. Start the TSL3000B and print the QR code for each bench. Paste the QR code to
each bench permanently.