Datasheet

DC voltage measurement accuracy
average mode
Average of 16 waveforms ±((DC Gain Accuracy) X |reading - (offset - position)| + Offset Accuracy + 0.11 div + 1 mV)
Delta Volts between any two
averages of ≥16 waveforms
acquired with the same
oscilloscope setup and
ambient conditions
±(DC Gain Accuracy X |reading| + 0.08 div + 1.4 mV)
Vertical position range ± 5 divisions
Vertical offset ranges
Volts/Div setting Offset range, 1 MΩ
2 mV/Div to 63 mV/Div ± 1 V
64 mV/div to 999 mV/div ± 10 V
1 V/div to 10 V/div ± 100 V
Analog bandwidth, DC coupled
200 MHz models: DC to >200 MHz
100 MHz models: DC to ≥100 MHz
70 MHz models: DC to ≥70 MHz
Common mode rejection ratio
(CMRR), typical
100:1 at 60 Hz, reducing to 10:1 with 50 MHz sine wave with equal Volts/div and coupling settings on each channel.
Channel-to-channel isolation
TBS2072B, TBS2074B
TBS2102B, TBS2104B TBS2202B, TBS2204B
≥100:1 at ≤70 MHz ≥100:1 at ≤100 MHz ≥100:1 at ≤200 MHz
Horizontal system analog channels
Maximum duration of time
captured at highest sample rate
(all channels)
5 ms
Time base range TBS220x: 1 ns/div to 100 sec/div
TBS207x, TBS210x: 2 ns/div to 100 sec/div
Time-base delay time range -15 divisions to 5000 s
Deskew range ±100 ns
Time base accuracy ±25 ppm over any ≥1 ms interval
Datasheet
Vertical system analog channels
10 www.tektronix.com