User Manual

Recalling data
Recalling data
The oscilloscope provides permanent internal memory locations from which you can recall instrument setups and waveforms.
You can also recall (load) setups and waveforms from les on an external USB drive.
Recalling oscilloscope setup information
Use this proc
edure to recall (load) oscilloscope settings from a m emory location or external le and set the oscilloscope to
those settings.
You can recall (load) saved s etup data to quickly congure the oscilloscope to take a certain measurement. A setup le
contains most of the oscilloscope settings, including vertical, horizontal, trigger, cursor, and measurement parameters. It
does not inc
lude communications information, such as GPIB or LAN conguration.
1. (Optional
) Install a USB drive to recall the setup data from a le on the USB drive.
2. Push the Sa
ve/Recall front-panel button.
3. Push the Action side-menu button.
4. Use the Multipurpose knob to select and click Recall Setup .
5. To recall setup data from internal setup memory:
a. Push the Recall From side-menu button.
b. Select and click Setup.
c. Push the Setup side-menu button and use the Multipurpose knob to select the setup memory location (1-10).
d. Push the Recall side-menu button. The oscilloscope changes its settings to those that are in the specied setup
memory.
6. To recall (load) setup data from a le on the USB drive:
a. Push the Recall From side-menu button.
b. Select and click USB File (*.SET).
c. Push the Select File side-menu button.
d. Highlight the setup le (*.SET) to load.
e. Push the Multipurpose knob or the Recall side-menu button. The oscilloscope loads and implements the settings.
TBS2000 Series User Manual 71