User Manual
Recalling data
Recalling data
The oscilloscope provides permanent internal memory locations from which you can recall instrument setups and waveforms.
You can also recall (load) setups and waveforms from files on an external USB drive.
Recalling oscilloscope setup information
Use this proc
edure to recall (load) oscilloscope settings from a m emory location or external file and set the oscilloscope to
those settings.
You can recall (load) saved s etup data to quickly configure the oscilloscope to take a certain measurement. A setup file
contains most of the oscilloscope settings, including vertical, horizontal, trigger, cursor, and measurement parameters. It
does not inc
lude communications information, such as GPIB or LAN configuration.
1. (Optional
) Install a USB drive to recall the setup data from a file on the USB drive.
2. Push the Sa
ve/Recall front-panel button.
3. Push the Action side-menu button.
4. Use the Multipurpose knob to select and click Recall Setup .
5. To recall setup data from internal setup memory:
a. Push the Recall From side-menu button.
b. Select and click Setup.
c. Push the Setup side-menu button and use the Multipurpose knob to select the setup memory location (1-10).
d. Push the Recall side-menu button. The oscilloscope changes its settings to those that are in the specified setup
memory.
6. To recall (load) setup data from a file on the USB drive:
a. Push the Recall From side-menu button.
b. Select and click USB File (*.SET).
c. Push the Select File side-menu button.
d. Highlight the setup file (*.SET) to load.
e. Push the Multipurpose knob or the Recall side-menu button. The oscilloscope loads and implements the settings.
TBS2000 Series User Manual 71