User Manual

Acquisition set
up
Maximum Zoo
med waveform record cap tured with 2 million points
Using the r
oll display mode
Roll mode gives a display similar to a strip chart recorder, in that the waveform moves slowly, or rolls, from right to left on
the screen. Roll mode is used to display low-frequency signals. Roll mode lets you s ee acquired data points without
waiting for the acquisition of a complete waveform record.
Roll mode is not a button- or menu-selected mode, but is present when the trigger mode is Auto and the horizontal scale and
record lengths are set as follows:
Table 1: Roll mode is enabled when:
Horizontal scale setting Record length (number of samples)
40 ms/div
2000, 20k, 200k, 2M
400 ms/div
20M
Roll mode tips
Push Run/Stop to halt Roll mode.
Role mode is disabled under the following conditions:
When using Reference, Math,orFFT waveforms
Using the Zoom function
Taking measurements (Measure menu)
Change to Normal trigger mode
Set the horizontal scale to 20 ms/division or faster (for 20M record lengths, it is 200 ms/division or faster)
Average acquisition mode is not available w hen you are using Roll mode