User Manual

Acquisition set
up
5. Push the Set Horizontal Position to 0s side-menu button to return the trigger point to the center of the waveform record
(setting delay to 0). Pushing this button does not turn off the delay mode.
You can also pu
sh the front-panel Horizontal Position knob to return the trigger point to the center of the w aveform record.
Trigger Delay is different than Trigger Holdoff. (See page 29, Holdoff trigger mode.)
Setting the r
ecord length
Use this procedure to set the record length. The record length sets how many samples (data points) are added to a waveform
record. Available record lengths are 2000; 20,000; 200,000; 2 m illion (2M); and 20 million (20M) points.
NOTE. 20M waveforms c annot be loaded into reference memory, either directly from the oscilloscope or from a saved le.
All other record length waveforms can be loaded into reference memory.
1. Push Acquire.
2. Push the
Record Length side-menu button.
3. Use the Multipurpose knob to select and set the record length.
Use longer record lengths to either capture
more sa
mples for a w aveform record, or
capture more c ycles of a waveform, then
use the Zoom controls to investigate the
wavefo
rm in more detail.
Maximum Zoomed wave form rec ord ca ptu red w ith 200 0 points
44 TBS2000 Series User Manual