User Manual

Trigger setup
Triggering on a specied pulse width
Use this procedure to set the oscilloscope to trigger on a specied signal pulse condition. You can trigger when a pulse is
less than, greater than, equal to, or not equal to a set time perio d (width). The minimum pulse width s etting is 1 ns. (See
page 27, Available trigger types.)
Pulse width triggering is most often used to troubleshoot or analyze digital signals.
1. Push the Trigger Menu front-panel button.
2. Push the Type side-menu button.
3. Use the Mul
tipurpose knob to select and click Pu lse Width .
4. Push the Source side-menu button and use the Multipurpose knob to select and click the channel to use as the trigger
signal so
urce, or select AC Line to use the AC power source frequency as the trigger signal.
5. Push the Polarity side-menu button and use the Multipurpose knob to select and click the pulse polarity (positive
or negat
ive) to trigger on.
6. Push the Threshold side-menu button and use the Multipurpose knob to select and click the trigger level entry method:
Click t
he threshold eld and use the Multipurpose knob to manually set the trigger level.
Click Choose Preset and use the Multipurpose knob to set the trigger level to a predened signal logic family level
(TTL,
CMOS, ECL, and so on) or to 0.
Click Set to 50% to set the trigger level to 50% of the signal peak-to-peak level.
7. Push
the Trigger When side-menu button and use the Multipurpose knob to select and click the pulse width condition on
which to trigger. Then use the Multipurpose knob to enter the pulse width time parameter.
TBS2000 Series User Manual 37