User Manual
Operating requi
rements
Operating req
uirements
Make sure to operate the instrument within the following environmental and power requirements.
Environment requirements
Characteristic Description
Operating temperature 0 °C to +50 °C, with 5 °C/minute maximum gradient, noncondensing, up to 3000 meter
altitude
Operating humidity 5% to 95% relative humidity (% RH) up to +30 °C
5% to 60% RH above +30 °C up to +50 °C, noncondensing
Operating altitude Up to 3000 meters (9842 feet)
Power requirements
Characteristic Description
Power source voltage
100 V
AC
– 240 V
AC
±10% RMS, single phase
Power source frequency 50/60 Hz over entire source voltage range
400 Hz (360 Hz to 440 Hz) for 115 V
AC
(100 V
AC
– 132 V
AC
) RMS source voltage range
Power consumption All models: 80 W maximum
CAUTION. A ground connection through the power cord grounding conductor is essential for safe operation.
TBS2000 Series User Manual 5