User Manual

Important safet
y information
Beware of high voltages. Understand the voltage ratings for the probe you are using and do not exceed those ratings.
Two ratings are
important to know and understand:
The maximum me
asurement voltage from the probe tip to the probe reference lead
The maximum o
ating voltage from the probe reference lead to earth ground
These two volt
age ratings depend on the probe and your application. Refer to the Specications manual for more information.
WARNING. To prevent electrical shock, do not exceed the maximum measurement or maximum oating voltage for the
oscilloscope input BNC connector, probe tip, or probe reference lead.
Connect and disconn
ect properly.
Connect the probe output to the measurement product before connecting the
probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input.
Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from
the measurement pr
oduct.
Connect and disconnect properly. De-energize the circuit under test before connecting or disconnecting the current
probe.
Connect the probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages above the current probe voltage rating.
Inspect the probe and accessories. Before each use, inspect probe and accessories for damage (cuts, tears, or
defects in the probe body, accessories, or cable jacket). Do not use if damaged.
Service safety
summary
The Service safety summary section contains additional inform ation required to safely perform service on the product. Only
qualied personnel should perform service procedures. Read this Service safety summary and the General safety summary
before performing any service procedures.
To avoid elec
tric shock.
Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of
rendering rst aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains
power before removing any covers or panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect
power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing
components.
Verify safety a fter repair. Always recheck ground continuity and mains dielectric strength after performing a repair.
TBS2000 Series User Manual vii