Datasheet
Peak Detect Captures glitches as narrow as 4 nsec at all sweep speeds.
Average From 2 to 256 waveforms included in average.
Hi-Resolution Averages multiple sample of one acquisition interval into one waveform point.
Roll Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div
Horizontal system - Analog channels
Timebase accuracy
±25 x 10
-6
over any >1 ms interval
Timebase range
TBS1202C,
TBS1052C,
TBS1072C, TBS1
102C
2 ns/div to 100 sec/div in a 1-2-4 sequence
Horizontal zoom Horizontally expand or compress a live or stopped waveform
Deskew range ± 100 nsec
Trigger system
External trigger input Included on all models
Trigger modes Auto, Normal, Single Sequence
Trigger Types
Edge Positive or negative slope on any channel. Coupling includes DC, HF reject, LF reject, and noise reject.
Pulse Width Trigger on width of positive or negative pulses that are >, <, =, or ≠ a specified period of time.
Runt Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.
Trigger source CH1, CH2, AUX IN, AC Line
Trigger Coupling DC, Noise Reject, High Frequency Reject, Low Frequency Reject
Trigger signal frequency readout Provides a frequency readout of the trigger source up to instrument bandwidth.
TBS1000C Series Oscilloscope Datasheet
www.tek.com 9