Datasheet
Waveform measurements
Cursors Time, Amplitude, Screen
Automated measurements 32, of which up to six can be displayed on-screen at any one time. Measurements include: Period, Frequency, Rise Time,
Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase,
Positive Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle
RMS, Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area, Cycle Area, Delay FR,
Delay FF, Delay FR, and Delay RR.
Gating Isolate the specific occurrence within an acquisition to take measurements on, using either the screen, between waveform
cursors or full record length.
Waveform math
Arithmetic Add, Subtract, and Multiply waveforms
FFT Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming,
Hanning, or Blackman-Harris.
Display system
Display Type 7-inch TFT Color Display
Display Resolution 800 horizontal by 480 vertical displayed pixels (WVGA)
Waveform styles Vectors, Variable persistence, and Infinite persistence.
Format YT and XY
Input Output Ports
USB 2.0 Host Port Supports USB mass storage devices
USB 2.0 device port Rear-panel connector allows for communication/control of oscilloscope through USBTMC or GPIB with a TEK-USB-488
Probe Compensator
TBS1000C Series Oscilloscope Datasheet
www.tek.com 10