User manual

Appendix A: Specications
Table 2: Input Specications (cont.)
Characteris
tic
Description
Maximum Input Voltage
At front panel connector, 300 V RMS, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and
above.
Based upon sinusoidal or DC input signal. Maximum viewable signal while DC coupled is ±50 V offset ± 5 V/division at 4 divisions, or
70 V. AC coupling allows measuring signals on a DC level up to 300 V. For non-sinusoidal waveforms, peak value must be less than
450 V. Excu
rsion above 300 V should be less than 100 ms duration and the duty factor is limited to < 44%. RMS signal level must be
limited to 300 V. If these values are exceeded, damage to the instrument may result.
Common Mod
e Rejection
Ratio (CMRR), typical
With the sa
me signal applied to each channel, CMRR is the ratio of the acquired signal amplitude to the amplitude of the MATH
difference waveform, either (Ch1 - Ch2), (Ch2 - Ch1), (Ch3 - Ch4) or (Ch4 - Ch3).
TBS1154, TBS1152, TBS1104, TBS1102: 100:1 at 60 Hz, reducing to 10:1 with 50 MHz sine wave, with equal Volts/division and
Coupling
settings on each channel.
TBS1064, TBS1062, TBS1042, TBS1022: 100:1 at 60 Hz, reducing to 20:1 with a sine wave with frequency equal to ½ the -3 dB
bandwidth and with equal Volts/division and Coupling settings on each channel.
Crosstalk (Channel Isolation) The ratio of the level of a signal input into one channel to that of the same signal present in another channel due to stray coupling.
TBS1154, TBS1152: 100:1 with 75 MHz sine wave and with equal V/div settings on each channel.
TBS1104
, TBS1102: 100:1 with 50 MHz sine wave and with equal V/div settings on each channel.
TBS1064, TBS1062: 100:1 with 30 MHz sine wave and with equal V/div settings on each channel.
TBS1042: 100:1 with 20 MHz sine wave and with equal V/div settings on each channel.
TBS102
2: 100:1 with 10 MHz sine wave and with equal V/div settings on each channel.
Table 3: Vertical Specications
Chara
cteristic
Descr
iption
Number of Digitized Bits 8 bits except at 2 mV/div
The nu
mber of bits in each binary word that results from the A/D converting each analog value value to a digitized value. (IEEE
standard 1057, Section 2.2.1)
Displayed vertically with 25 digitalization levels per division, 10 divisions dynamic range.
2mV/
division setting is generated by digital multiplication and the resolution is reduced. Given 100 levels available, the resolution is
>6.5 bits.
Sen
sitivity Range
2mV
/Div to 5 V/Div in 1-2-5 sequence with probe attenuation set to 1X
The position ranges are as follows
Volts/Div Setting
Pos
ition Range
2m
V/div to 200 mV/div
±1.8 V
Vertical Position Ranges
>200 mV/div to 5 V/div
±45 V
Analog Bandwidth, DC
Coupled, Sample or
A
verage
V/div values are accurate for probe attenuation settings of 1X. No probe should be installed for these measurements.
TBS1154, TBS1152: DC to 150 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. <5 mV/div settings are
l
imited to 20 MHz bandwidth.
TBS1104, TBS1102: DC to 100 MHz for 5 mV through 5 V/div settings with bandwidth limit at full. <5 mV/div settings are limited to
20 MHz bandwidth.
T
BS1064, TBS1062: DC to 60 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. <5 mV/div settings are
limited to 20 MHz bandwidth.
TBS1042: DC to 40 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. <5 mV/div settings are limited to
20 MHz bandwidth.
TBS1022: DC to 25 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. <5 mV/div settings are limited to
20 MHz bandwidth.
108 TBS1000 Series Oscilloscopes User Manual