Instructions

6. Disconnect the test setup.
7. Repeat steps 1 on page 15 through 6 on page 16 for all input channels.
Check bandwidth
This test checks the bandwidth of all input channels.
1. Set up the oscilloscope using the following table:
Push menu button Select menu option Select setting
Default Setup - -
Channel 1 Probe 1X
Acquire Average 16
Trig Menu Coupling Noise Reject
Measure Source Channel under test
Measurements Peak-Peak
2. Connect the oscilloscope channel under test to the leveled sine wave generator as shown in the following figure:
3. Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
4. Set the oscilloscope Horizontal Scale (seconds/division) to 10 μs/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-peak measurement is between 2.98 V and 3.02 V.
7. Set the leveled sine wave generator frequency to:
200 MHz if you are checking a TBS1202C
100 MHz if you are checking a TBS1102C
70 MHz if you are checking a TBS1072C
50 MHz if you are checking a TBS1052C
8. Set the oscilloscope Horizontal Scale (seconds/division) to 10 ns/div.
9. Check that the peak-to-peak measurement is ≥ 2.12 V.
10. Disconnect the test setup.
11. Repeat steps 1 on page 16 through 10 on page 16 for all input channels.
Check sample rate accuracy and delay time accuracy
Performance verification
TBS1000 Series Oscilloscopes Series Digital Storage Oscilloscope Technical Reference 16