Instructions

Display specifications
Display type 7 inch TFT Liquid Crystal Display (LCD).
Display Resolution This is the number of individually addressable pixels
800 horizontal by 480 vertical displayed pixels
Format YT and XY
Waveform styles Vectors, variable persistence, and Infinite persistence.
Luminance, typical
300 cd/m
2
, typical.
Contrast ratio and control, typical Available black room contrast ratio, full black to full white. 400 minimum, 500 typical.
Interfaces and output ports specifications
USB Interface 1 High Speed USB2.0 Host and 1 High Speed USB2.0 Device connector are
standard in all models.
Kensington style lock Rear-Panel security slot connects to standard Kensington-style lock
Probe compensator, output
voltage and frequency, typical
The Probe compensator output voltage is in peak-to-peak volts and frequency is in
Hertz.
0-5V amplitude +/- 10%, 1 kHz +/- 10%
Data handling characteristics
Retention of front panel settings Front panel settings are stored periodically in memory. The settings are not lost when
the instrument is turned off or if there is a power failure.
Stored waveforms and multiple
front panel settings
Two Channel 1, Channel 2, or Math waveforms can be stored in nonvolatile
waveform memory A or B. One, both, or neither of A or B waveform memories can
be displayed. Ten user setups of the current instrument settings can be saved and
restored from nonvolatile memory. Additional storage is available when an
appropriate mass storage device is connected via USB.
Power distribution system
Power consumption 30 W (typ)
Source voltage Full range: 100 to 240 VACRMS ± 10%, Installation Category II (covers range of 90
to 264 V
AC
)
Power source 100 to 240 VAC RMS ±10%
Power source frequency 45 Hz to 65 Hz (100 to 240 V).
360 Hz to 440 Hz (100 to 132 V).
Fuse rating 3.15 Amps, T rating, 250 V; IEC and UL approved.
Specifications
TBS1000 Series Oscilloscopes Series Digital Storage Oscilloscope Technical Reference 10