Datasheet
Specifications
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.
Model overview
Parameters TBS1052C TBS1072C TBS1102C TBS1202C
Bandwidth 50 MHz 70 MHz 100 MHz 200 MHz
Channels 2 2 2 2
Sample Rate 1 GS/s 1 GS/s 1 GS/s 1 GS/s
Record Length 20 K points 20 K points 20 K points 20 K points
Vertical system - Analog channels
Vertical resolution 8 bits
Input sensitivity range 1 mV/div to 10 V/div max. in 1-2-5 sequence with probe attenuation set to 1X
DC gain accuracy
±3.0 % step gain, derated at 0.1 %/°C above 30 °C
Maximum Input Voltage 300 VRMS, Installation Category II; derate above 4 MHz at 20 dB per decade to 200 MHz
Offset range 1 mV/div to 50 mV/div: ± 1 V
100 mV/div to 500 mV/div: ± 10 V
1 V/div to 5 V/div : ± 100 V
Bandwidth limit 20 MHz (Typ)
Input coupling DC, AC
Input impedance 1 MΩ ±2 % in parallel with 14 pF ±2 pF
Vertical zoom Vertically expand or compress a live or stopped waveform
Acquisition modes
Sample Acquire sampled values
Peak Detect Captures glitches as narrow as 4 nsec at all sweep speeds.
Average From 2 to 256 waveforms included in average.
Hi-Resolution Averages multiple sample of one acquisition interval into one waveform point.
Roll Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div
www.tek.com 7