Datasheet

Color palettes Normal and inverted for screen captures
Individual waveform colors are user-selectable
Format YT, XY, and XYZ
Local Language User Interface
and Help
English, Japanese, Simplified Chinese
Arbitrary-Function Generator (optional)
Function types Arbitrary, sine, square, pulse, ramp, triangle, DC level, Gaussian, Lorentz, exponential rise/fall, sin(x)/x, random noise, Haversine,
Cardiac
Sine waveform
Frequency range 0.1 Hz to 50 MHz
Frequency setting resolution 0.1 Hz
Frequency accuracy 130 ppm (frequency ≤ 10 kHz), 50 ppm (frequency > 10 kHz)
This is for Sine, Ramp, Square and Pulse waveforms only.
Amplitude range 20 mV
pp
to 5 V
pp
into Hi-Z; 10 mV
pp
to 2.5 V
pp
into 50 Ω
Amplitude flatness, typical ±0.5 dB at 1 kHz
±1.5 dB at 1 kHz for < 20 mV
pp
amplitudes
Total harmonic distortion,
typical
1% for amplitude ≥ 200 mV
pp
into 50 Ω load
2.5% for amplitude > 50 mV AND < 200 mV
pp
into 50 Ω load
This is for Sine wave only.
Spurious free dynamic range,
typical
40 dB (V
pp
≥ 0.1 V); 30 dB (V
pp
≥ 0.02 V), 50 Ω load
Square and pulse waveform
Frequency range 0.1 Hz to 25 MHz
Frequency setting resolution 0.1 Hz
Duty cycle range 10% - 90% or 10 ns minimum pulse, whichever is larger
Minimum pulse time applies to both on and off time, so maximum duty cycle will reduce at higher frequencies to maintain 10 ns off
time
Duty cycle resolution 0.1%
Minimum pulse width, typical 10 ns. This is the minimum time for either on or off duration.
Rise/Fall time, typical 5.5 ns, 10% - 90%
Pulse width resolution 100 ps
Overshoot, typical < 4% for signal steps greater than 100 mV
pp
This applies to overshoot of the positive-going transition (+overshoot) and of the negative-going (-overshoot) transition
Asymmetry, typical ±1% ±5 ns, at 50% duty cycle
Jitter, typical < 60 ps TIE
RMS
, ≥ 100 mV
pp
amplitude, 40%-60% duty cycle
Ramp and triangle waveform
Frequency range 0.1 Hz to 500 kHz
Frequency setting resolution 0.1 Hz
Variable symmetry 0% - 100%
Symmetry resolution 0.1%
4 Series MSO