Datasheet

Digital volt meter (DVM)
Measurement types DC, AC
RMS
+DC, AC
RMS
Voltage resolution 4 digits
Voltage accuracy
DC: ±((1.5% * |reading - offset - position|) + (0.5% * |(offset - position)|) + (0.1 * Volts/div))
De-rated at 0.100%/°C of |reading - offset - position| above 30 °C
Signal ± 5 divisions from screen center
AC: ± 2% (40 Hz to 1 kHz) with no harmonic content outside 40 Hz to 1 kHz
AC, typical: ± 2% (20 Hz to 10 kHz)
For AC measurements, the input channel vertical settings must allow the V
PP
input signal to cover between 4 and 10 divisions and
must be fully visible on the screen
Trigger frequency counter
Accuracy ±(1 count + time base accuracy * input frequency)
The signal must be at least 8 mV
pp
or 2 div, whichever is greater.
Maximum input frequency 10 Hz to maximum bandwidth of the analog channel
The signal must be at least 8 mV
pp
or 2 div, whichever is greater.
Resolution 8-digits
Processor system
Host processor ARM 1.5 GHz, 32-bit, dual core processor
Internal storage 64 GB eMMC
Input-Output ports
HDMI-Video-Port
HDMI video port A 29-pin HDMI connector
Supported resolution: 1920 x 1080 @ 60Hz (only). The monitor must be attached before powering on the instrument
Probe compensator signal, typical
Connection: Connectors are located on the lower right-hand side of the instrument
Amplitude: 0 to 2.5 V
Frequency: 1 kHz
Source impedance: 1 kΩ
External reference input The time-base system can phase lock to an external 10 MHz reference signal (±4 ppm).
USB interface (Host, Device ports) Front panel USB Host ports: Three USB 2.0 Hi-Speed ports
Rear panel USB Host ports: Two USB 2.0 Hi-Speed ports
Rear panel USB Device port: One USB 2.0 High Speed Device port providing USBTMC support
Ethernet interface 10/100/1000 Mb/s
4 Series MSO
www.tek.com 29