Datasheet

Vertical system - digital channels
Number of channels 8 digital inputs (D7-D0) per installed TLP058 (traded off for one analog channel)
Vertical resolution 1 bit
Minimum detectable pulse width,
typical
1 ns
Thresholds One threshold per digital channel
Threshold range ±40 V
Threshold resolution 10 mV
Threshold accuracy ± [100 mV + 3% of threshold setting after calibration]
Input hysteresis, typical 100 mV at the probe tip
Input dynamic range, typical 30 V
pp
for F
in
≤ 200 MHz, 10 V
pp
for F
in
> 200 MHz
Absolute maximum input voltage,
typical
±42 V peak
Minimum voltage swing, typical 400 mV peak-to-peak
Input impedance, typical 100 kΩ
Probe loading, typical 2 pF
Horizontal system
Time base range 200 ps/div to 1,000 s/div
Sample rate range 1.5625 S/s to 6.25 GS/s (real time)
12.5 GS/s to 500 GS/s (interpolated)
Record length range
Standard 1 kpoints to 31.25 Mpoints in single sample increments
Option 4-RL-1 62.5 Mpoints
Maximum duration at highest
sample rate
5 ms (std.) or 10 ms (opt.)
Time base delay time range -10 divisions to 5,000 s
Deskew range -125 ns to +125 ns with a resolution of 40 ps
Timebase accuracy ±2.5 x 10
-6
over any ≥1 ms time interval
Description
Specification
Factory Tolerance ±5.0 x10
-7
. At calibration, 25 °C ambient, over any ≥1 ms interval
Temperature stability, typical ±5.0 x10
-7
. Tested at operating temperatures
Crystal aging ±1.5 x 10
-6
. Frequency tolerance change at 25 °C over a period of 1 year
4 Series MSO
www.tek.com 21