Datasheet

Specifications
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.
Model overview
Oscilloscope
MSO44 MSO46
FlexChannel inputs 4 6
Maximum analog channels 4 6
Maximum digital channels (with
optional logic probes)
32 48
Auxiliary Trigger Input ≤300 V
RMS
(Edge Trigger only)
Bandwidth (calculated rise time) 200 MHz, 350 MHz, 500 MHz, 1 GHz, 1.5 GHz
DC Gain Accuracy 50 Ω: ±1%, (±2.5% at 1 mV/Div and 500 µV/Div settings), de-rated at 0.100%/°C above 30 °C
1 MΩ and 250 kΩ: ±1.0%, (±2.0% at 1 mV/Div and 500 µV/Div settings)
ADC Resolution 12 bits
Vertical Resolution 8 bits @ 6.25 GS/s
12 bits @ 3.125 GS/s
13 bits @ 1.25 GS/s (High Res)
14 bits @ 625 MS/s (High Res)
15 bits @ 312.5 MS/s (High Res)
16 bits @ ≤125 MS/s (High Res)
Sample Rate 6.25 GS/s on all analog / digital channels (160 ps resolution)
Record Length (std.) 31.25 Mpoints on all analog / digital channels
Record Length (opt.) 62.5 Mpoints on all analog / digital channels
Waveform Capture Rate, typical >500,000 wfms/s
Arbitrary/Function Generator (opt.) 13 predefined waveform types with up to 50 MHz output
DVM 4-digit DVM (free with product registration)
Trigger Frequency Counter 8-digit frequency counter (free with product registration)
Vertical system - analog channels
Bandwidth selections 50 Ω: 20 MHz, 250 MHz, and the full bandwidth value of your model
1 MΩ: 20 MHz, 250 MHz, 500 MHz
Input coupling DC, AC
Input impedance 50 Ω ± 1%
1 MΩ ± 1% with 13.0 pF ± 1.5 pF
Input sensitivity range
1 MΩ 500 µV/div to 10 V/div in a 1-2-5 sequence
50 Ω 500 µV/div to 1 V/div in a 1-2-5 sequence
Note: 500 μV/div is a 2X digital zoom of 1 mV/div or a 4x zoom of 2 mV/div depending upon instrument settings
4 Series MSO
www.tek.com 19