Datasheet

Lorentz
Frequency range (typical) 0.1 Hz to 5 MHz
Amplitude range 20 mV
p-p
to 2.4 V
p-p
into Hi-Z; 10 mV
p-p
to 1.2 V
p-p
into 50 Ω
Exponential Rise / Decay
Frequency range (typical) 0.1 Hz to 5 MHz
Amplitude range 20 mV
p-p
to 2.5 V
p-p
into Hi-Z; 10 mV
p-p
to 1.25 V
p-p
into 50 Ω
Haversine
Frequency range (typical) 0.1 Hz to 5 MHz
Amplitude range 20 mV
p-p
to 2.5 V
p-p
into Hi-Z; 10 mV
p-p
to 1.25 V
p-p
into 50 Ω
Cardiac (typical)
Frequency range 0.1 Hz to 500 kHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Arbitrary
Memory depth 1 to 128 k
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Repetition rate 0.1 Hz to 25 MHz
Sample rate 250 MS/s
Frequency accuracy
Sine wave and ramp 130 ppm (frequency < 10 kHz)
50 ppm (frequency ≥ 10 kHz)
Square wave and pulse 130 ppm (frequency < 10 kHz)
50 ppm (frequency ≥ 10 kHz)
Resolution 0.1 Hz or 4 digits; whichever is larger
Signal amplitude accuracy ±[ (1.5% of peak-to-peak amplitude setting) + (1.5% of DC offset setting) + 1 mV ] (frequency = 1 kHz)
DC offset
DC offset range ±2.5 V into Hi-Z;
±1.25 V into 50 Ω
DC offset resolution 1 mV into Hi-Z;
500 uV into 50 Ω
DC offset accuracy ±[(1.5% of absolute offset voltage setting) + 1 mV] Add 3 mV for every 10 °C change from 25 °C
3 Series MDO
www.tek.com 23