Datasheet

Act on Event
Events None, when a trigger occurs, or when a defined number of acquisitions complete (1 to 1,000,000)
Actions Stop acquisition, save waveform to file, save screen image, print, AUX OUT pulse, remote interface SRQ, e-mail notification, and
visual notification
Repeat Repeat the act on event process (1 to 1,000,000 and infinity)
Power measurements (optional)
Power quality measurements V
RMS
, V
Crest Factor
, Frequency, I
RMS
, I
Crest Factor
, True Power, Apparent Power, Reactive Power, Power Factor, Phase Angle.
Switching loss measurements
Power loss T
on
, T
off
, Conduction, Total.
Energy loss T
on
, T
off
, Conduction, Total.
Harmonics THD-F, THD-R, RMS measurements. Graphical and table displays of harmonics. Test to IEC61000-3-2 Class A and MIL-
STD-1399, Section 300A.
Ripple measurements V
Ripple
and I
Ripple
.
Modulation analysis Graphical display of +Pulse Width, -Pulse Width, Period, Frequency, +Duty Cycle, and -Duty Cycle modulation types.
Safe operating area Graphical display and mask testing of switching device safe operating area measurements.
dV/dt and dI/dt measurements Cursor measurements of slew rate
Spectrum Analyzer
(Requires 3-SA1 or 3-SA3 option)
Capture bandwidth All models: 1 GHz with option 3-SA1 or 3 GHz with option 3-SA
Span All models: 9 kHz - 1 GHz with option 3-SA1 or 3 GHz with option 3-SA3, in a 1-2-5 sequence
Resolution bandwidth 20 Hz - 150 MHz in a 1-2-3-5 sequence
Reference level -140 dBm to +20 dBm in steps of 5 dBm
Vertical scale 1 dB/div to 20 dB/div in a 1-2-5 sequence
Vertical position -100 divs to +100 divs (displayed in dB)
Vertical units dBm, dBmV, dBµV, dBµW, dBmA, dBµA
Displayed average noise level
(DANL)
9 kHz - 50 kHz < -109 dBm/Hz (< -113 dBm/Hz typical)
50 kHz – 5 MHz < -126 dBm/Hz (< -130 dBm/Hz typical)
5 MHz - 2 GHz < -136 dBm/Hz (< -140 dBm/Hz typical)
2 GHz – 3 GHz < -126 dBm/Hz (< -130 dBm/Hz typical)
3 Series MDO