User's Manual
Performance Tests
AWG2005 Service Manual
4Ć27
3. Disconnect connections to the test
equipment.
These procedures check the accuracy of the AWG2005 clock frequency and
the waveform output amplitude.
Check Clock Frequency Accuracy
Electrical Characteristic Checked: Clock Generator, Accuracy, on
pageĂ1Ć11.
Equipment Required: A 50ĂΩ coaxial cable and a frequency counter.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1.
a. Connect the AWG2005 rear panel
CLOCK OUT connector to the frequency counter input through a
coaxial cable (see Figure 4Ć11).
AWG2005
AWG2005
Frequency Counter
DC 5010
Figure 4Ć11:ăClock Frequency Accuracy Initial Test Hookup
Clock Frequency and
Amplitude Checks