User's Manual
!
!
4Ć26
b.
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 0ĂV
Offset 5.0ĂVĂ
(The actual voltage out is
1
/
2
of what is displayed.)
Output Off
2. Turn on the function generator
output.
3.
a.
Push the AWG2005 button, and check that
the step changes to Step 4 on the MODE menu.
Check that the DMM reading is in the range from 4.750 to
5.250ĂV.
4. Turn the function generator output off.
5. Keep the test connections and instrument settings for
the next check.
! !
! ! ! Arithmetic Operation, Add, on
pageĂ1Ć13.
"! " Two 50ĂΩ coaxial cables, a function generator, and
a digital multimeter (DMM).
" ! The AWG2005 meets the prerequisites listed on page 4Ć7.
"
1.
2.
a.
Push the AWG2005 button, and check that
the step changes to Step 5 on the MODE menu.
Check that the DMM reading is in the range from 4.750 to
5.250ĂV.