User's Manual

  !
 !
4Ć26
b.    
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 0ĂV
Offset 5.0ĂVĂ
(The actual voltage out is
1
/
2
of what is displayed.)
Output Off
2.      Turn on the function generator
output.
3.     
a.    
Push the AWG2005   button, and check that
the step changes to Step 4 on the MODE menu.
Check that the DMM reading is in the range from 4.750 to
5.250ĂV.
4.     Turn the function generator output off.
5.   Keep the test connections and instrument settings for
the next check.
 !  !
! ! !  Arithmetic Operation, Add, on
pageĂ1Ć13.
"! " Two 50Ăcoaxial cables, a function generator, and
a digital multimeter (DMM).
" !  The AWG2005 meets the prerequisites listed on page 4Ć7.
"
1.         
2.    
a.    
Push the AWG2005   button, and check that
the step changes to Step 5 on the MODE menu.
Check that the DMM reading is in the range from 4.750 to
5.250ĂV.