User's Manual
Performance Tests
AWG2005 Service Manual
4Ć21
OscilloscopeAWG2005
AWG2005
Figure 4Ć8:ăAutostep Mode Initial Test Hookup
b.
Vertical CH1
CH1 coupling DC
CH1 scale 0.5ĂV/div.
CH1 input impedance 50ĂΩ
Horizontal
Sweep 50õs/div.
Trigger
Source CH2
Coupling DC
Slope Positive
Level 100ĂmV
Mode Auto
2. !
a. # Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
b. "
Push MODEAutostepSlope to highlight Positive.
Push Run from the side menu to highlight Continuous.
Select Level from the side menu, and turn the general purpose
knob to select 1ĂV.
Select Select Autostep File from the side menu.
Turn the general purpose knob to highlight the MODE_AST.AST
file.
Push ENTER to select the file.
3. Push the AWG2005 MANUAL TRIGGER button
and check that the oscilloscope momentarily displays a sine wave with a
different frequency and amplitude each time you push the button.
4. Disconnect the oscilloscope.