User's Manual
Performance Tests
AWG2005 Service Manual
4Ć17
Horizontal
Sweep 200õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 500ĂmV
Mode Auto
d. Set function generator controls:
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 4.0 V
Offset 2.0 V
Output Off
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
b. Modify the AWG2005 default settings:
Push MODEGatedPolarity to highlight Positive.
c. Select the file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE.WFM file, using the general purpose knob.
Push ENTER to select the file.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check gated mode with manual trigger: Push and hold the AWG2005
MANUAL TRIGGER button, and check that the oscilloscope continuĆ
ously displays a sine wave while the MANUAL TRIGGER button is
pushed.
5. Check gated mode with gate signal:
a. Change the oscilloscope controls:
Vertical CH1
CH1 coupling DC
CH1 scale 0.5ĂV/div.
CH1 input impedance 50ĂΩ