User's Manual
Performance Tests
Performance Verification
4Ć16
6. Turn off the function generator output, and disconnect
the function generator and oscilloscope.
Check Gated Mode
Electrical Characteristic Checked: Operating modes, Gated, on
pageĂ1Ć5.
Equipment Required: Three 50ĂΩ coaxial cables, a 50 Ω precision terĆ
mination, a function generator, and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1.
a. Connect the AWG2005 CH1 output
connector through the coaxial cable to the CH1 vertical input conĆ
nector on the oscilloscope.
b. Connect the function generator
output to both the AWG2005 TRIGGER INPUT and the oscilloscope
CH2 input through a coaxial cable and a dual input coupler (see
Figure 4Ć5).
Function Generator
FG 5010
OscilloscopeAWG2005
AWG2005
Figure 4Ć5:ăGated Mode Initial Test Hookup
c.
Vertical CH1
CH1 coupling DC
CH1 scale 0.5ĂV/div.
CH1 input impedance 50ĂΩ