User's Manual

Performance Tests
AWG2005 Service Manual
4Ć15
Trigger
Source CH1
Coupling DC
Slope Positive
Level -100ĂmV
Mode Auto
d. Set the function generator controls:
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 4ĂV
Offset 2ĂV
Output Off
2. Set AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
b. Modify the AWG2005 default settings:
Push MODETriggeredSlope to select Positive slope.
Select Level from the side menu and turn the general purpose
knob to select a 1ĂV trigger level.
c. Select the file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE.WFM file using the general purpose knob.
Push ENTER to select the file.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check triggered mode with manual triggering: Push the AWG2005
MANUAL TRIGGER button and check that when the button is pushed,
the oscilloscope displays a oneĆcycle sine wave.
5. Check triggered mode with external triggering:
a. Change the oscilloscope trigger mode to Normal.
b. Enable function generator output: Turn on the function generator
output.
c. Check triggering: Check that for each trigger supplied by the funcĆ
tion generator, the oscilloscope displays a oneĆcycle sine wave.