User's Manual
Performance Tests
Performance Verification
4Ć42
6. Turn off the function generator output and disconnect
the function generator.
This procedure checks the AWG2005 response to an external CLOCK IN
signal.
Electrical Characteristic Checked: Auxiliary Input, CLOCK, Threshold
level, on pageĂ1Ć15.
Equipment Required: Two 50ĂΩ coaxial cables, a function generator, and
an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1.
a. Connect the AWG2005 CH1 output through a
coaxial cable to the oscilloscope CH1 vertical input.
b. Connect the AWG2005 rearĆpanel
CLOCK IN through a coaxial cable to the function generator output
(see Figure 4Ć19).
AWG2005
Oscilloscope
Function Generator
FG 5010
AWG2005
Figure 4Ć19:ăExternal CLOCK IN Initial Test Hookup
c.
External CLOCK IN
Check