User's Manual

Performance Tests
Performance Verification
4Ć40
OscilloscopeAWG2005
AWG2005
Function Generator
FG 5010
Figure 4Ć18:ăExternal Trigger Level Accuracy Initial Test Hookup
c.   
Vertical CH1
CH1 Coupling DC
CH1 Scale 0.2ĂV/div.
CH1 Input Impedance 50Ă
Horizontal
Sweep 50õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 0ĂV
Mode Auto
d.    
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 0ĂV
Offset 0.6ĂV
Output Off
2.         
a.     Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
b.    
Push MODEGatedPolarity to highlight Positive.